Model Number SIGTRSB60AMT
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Code Tissue Breakdown (2681)
Date of Event 03/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, in a laparoscopic gastric bypass sleeve, during the transection of the stomach, the powered stapler partially fired.The firing stop half way through the firing.The staple line was resected and re-stapled due to the issue.The issue happened with four reloads.
 
Additional Manufacturer Narrative
D10 concomitant products: sigadaptxl, sig power sigadaptxl linear xl adapter, (serial # (b)(6)); sigtrsb60amt, sigtrsb60amt 60mm med thk reinforced rel, (lot # n5l1164y); sigtrsb60amt, sigtrsb60amt 60mm med thk reinforced rel, (lot # n5l1164y); sigtrsb60amt, sigtrsb60amt 60mm med thk reinforced rel, (lot # n5l1164y); sigphandle, sig power sigphandle handle, (serial # unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 correction: e1(prefix should be blank) h3 evaluation summary: medtronic conducted an investigation based upon all information received.Four representative samples were available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device experienced partial firing.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24812907
Report Number1219930-2026-02296
Device Sequence Number6012103
Product Code GDW
UDI-Device Identifier10884521717626
UDI-Public(01)10884521717626(17)281031(10)N5L1164Y(20)04
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K173270
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSIGTRSB60AMT
Device Catalogue NumberSIGTRSB60AMT
Device Lot NumberN5L1164Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2026
Supplement Date Received by Manufacturer05/22/2026
Initial Report FDA Received Date04/08/2026
Supplement Report FDA Received Date06/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-