Product complaint # (b)(4).Date sent to the fda: 4/8/2026 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: did this issue contribute to any patient adverse event? no how many devices demonstrated the alleged deficiency? multiple sutures from multiple boxes broke a few throws into the anastomosis.Did the event occur during one or multiple patient procedures? multiple what is the total number of procedures? 5, with multiple surgeons.Have any of these events been previously reported to ethicon? if so, could you please provide the corresponding reference number(s)? they were reported to (please refer to the attached mail) when was the suture broke (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify the sutures broke wile in use on the patient.Please perform and document the follow up attempt for product return.Please provide tracking number unsure of the tracking information.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|