Catalog Number CXT201412
Medical Device Problem Code Device Endoleak (4074)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 03/11/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The following was reported to gore: on (b)(6) 2026, a patient was treated for an abdominal aortic aneurysm using a gore® excluder® conformable aaa endoprosthesis, as well as gore® excluder® aaa endoprosthesis devices.On an unknown date, a ct scan revealed a type 1a endoleak on the conformable trunk.On (b)(6) 2026, a reintervention occurred.A conformable extender was placed proximal to the cxt device, and 6 medtronic endoanchors were implanted.No leak was seen on pre-intervention or completion angiogram.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
H.6.Type of investigation code b15: an evaluation of pre-reintervention images confirmed the presence of the endoleak, but could not determine a root cause.H.6.Type of investigation code b20: the device remains implanted and is not available for analysis.H.6.Investigation findings code c19: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H.6.Investigation conclusions code d15: there is insufficient information available for gore to reasonably draw conclusions related to aspects of the event, therefore conclusion code ¿d15: cause not established¿ is being used.Insufficient information may include limited or missing relevant medical records, involvement of multiple implanted devices (including non-gore devices) in the field of treatment, patient non-compliance, and/or a general lack of available detail or specificity related to an adverse event and/or device.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
bryce wargin
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key24816511
Report Number3007284313-2026-04692
Device Sequence Number9594814
Product Code MIH
UDI-Device Identifier00733132650996
UDI-Public(01)00733132650996(17)280724
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P200030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue NumberCXT201412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2026
Initial Report FDA Received Date04/08/2026
Date Device Manufactured07/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight49 KG
-
-