• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 MP TYPE1 PPS HO 17.0; PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL UNCEMENTED ACETABULAR COMPONENT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 MP TYPE1 PPS HO 17.0; PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL UNCEMENTED ACETABULAR COMPONENT) Back to Search Results
Catalog Number 51-107170
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2: foreign, event occurred in (b)(6).Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility has been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during an internal stock inspection, a sterile package was found damaged.The affected item was identified in circulated inventory and not used clinically.There was no patient involvement.Attempts have been made and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP TYPE1 PPS HO 17.0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL UNCEMENTED ACETABULAR COMPONENT)
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key24817459
Report Number0001825034-2026-00877
Device Sequence Number15851588
Product Code KWA
UDI-Device Identifier00880304513570
UDI-Public(01)00880304513570(17)330621(10)7530444
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110400
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number51-107170
Device Lot Number7530444
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2026
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2026
Initial Report FDA Received Date04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-