| Model Number |
693565 |
| Medical Device Problem Codes |
Over-Sensing (1438); Defibrillation/Stimulation Problem (1573); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
Ventricular Fibrillation (2130)
|
| Date of Event |
03/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 439878 lead; implant date: (b)(6) 2018.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited oversensing in non-physiologic intervals and undersensing before ventricular fibrillation (vf) detection resulting in delay of episode detection.It was noted that the left ventricular (lv) lead was previously reprogrammed due to phrenic stimulation.The leads remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated right ventricular undersensing.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|