|
Additional information: d9, g3, h3, h6 the complaint allegation was confirmed.One unpackaged ar-9530xs-03 trial, humeral insert, 33 +3 batch number 465327 was received for evaluation.Visual inspection identified a device fracture, with additional material damage including dents and nicks.A functional test was not performed due to device damage.The most likely cause of the reported failure is use error, including excessive force applied to the device, with potential contribution from age-related degradation, manufactured in 2022.
|