| Model Number |
N/A |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A united states (us) customer contacted the siemens customer care center (ccc) regarding an erroneously elevated high-sensitivity troponin i (tnih) result on a patient sample obtained on the advia centaur cp instrument.Siemens evaluated the instrument data and the information provided by the customer.Reagent issues were ruled out based on review of quality control (qc) which showed recovery within acceptable ranges and no issues were reported with other patient samples.No observable clots, fibrin, or lipemia present in the patient sample.Based on the available information, the cause of the discordant result cannot be determined.A product problem has not been identified.The customer is operational, and the reported issue was resolved by routine troubleshooting.No further evaluation is required.
|
| |
|
Event or Problem Description
|
|
The customer reported an erroneously elevated high-sensitivity troponin i (tnih) result on a patient sample was obtained on an advia centaur cp instrument.The erroneous result was reported to the physician and was not questioned.A new sample was drawn from the same patient and tested on same advia centaur cp instrument.The result obtained was lower.Again, the initial sample was repeated on the original instrument and similar lower result was obtained.Both the results were same and considered correct.This repeat result was reported to the physician.There are no known reports of adverse health consequences or patient interventions due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens filed the initial mdr (1219913-2026-00072) on 08-apr-2026.Correction: previously in section d4, the primary udi number field was updated incorrectly.Therefore, the section d4 has been revised with the correct udi details.
|
| |
|
Search Alerts/Recalls
|