SMITH & NEPHEW, INC. 7.0 COMPRESSION SCREW DRILL; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71674070
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a trauma surgery, the tip of one (1) 7.0 compression screw drill broke off while in use, all the broken pieces were removed and accounted.The procedure was completed, after a non-significant delay, with the same s+n product.No further complications were reported.
 
Additional Manufacturer Narrative
H11: internal complaint reference case-(b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference:(b)(4).Additional information: h11: the associated device was returned and evaluated.The visual inspection revealed the device returned heavily worn from use, with the distal tip fractured off and returned with the device.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed part number over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7.0 COMPRESSION SCREW DRILL
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24820851
Report Number1020279-2026-00489
Device Sequence Number9630925
Product Code HSB
UDI-Device Identifier03596010560636
UDI-Public3596010560636
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71674070
Device Lot Number23BNG0042
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2026
Supplement Date Received by Manufacturer05/05/2026
Initial Report FDA Received Date04/08/2026
Supplement Report FDA Received Date05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/14/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-