OLYMPUS WINTER & IBE GMBH VIDEO TELESCOPE "ENDOEYE", 5 MM, 30°, NTSC, AUTOCLAVABLE, SHORT; RIGID VIDEO LAPAROSCOPE
|
Back to Search Results |
|
| Model Number |
WA50023S |
| Medical Device Problem Codes |
Break (1069); Poor Quality Image (1408); Dent in Material (2526); Detachment of Device or Device Component (2907); Environmental Particulates (2930); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the rigid video laparoscope had fiber breakage, dent on distal edge, scratches on distal edge, reflections on image, missing light guide adapter, and excessive dirt of objective lens.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure., expected or random component failure without any design or manufacturing issue due to degradation problem.Problems that occur when the device becomes worn, weakened, corroded, or broken down due to processes such as aging, permeation, and corrosion.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|