| Model Number |
SIF-Q180 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation.The most probable cause of the complaint is cause not established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the small intestinal videoscope exhibited white residue in air water channel and cylinder.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|