ETHICON ENDO-SURGERY, LLC. N/A; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECS29A
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 4/8/2026.D4: batch # a98v6k.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.Additional information was requested and the following was obtained: 1.Please provide more details for "open is not exciting".The surgeon noted the device could not fire after open the packaging.2.Was the device locked on tissue with the anvil closed? unknown.4.Did the surgeon turn the rotation knob full revolutions as stated in the ifu? unknown.5.If the anvil could not be opened, how was the device removed.Unknown.6.Was the device not able to be removed and subsequently the tissue was cut? unknown.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the unk surgery, could not fire.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 4/24/2026.D4: batch # 756d12.Investigation summary: the product was returned to ethicon endo-surgery for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample revealed that the ecs29a device arrived with no apparent damage along with the stapler retainer loose.The breakaway washer was present and uncut, the device was fully loaded with staples, indicating that the device had not been fired.The device was tested for functionality with the returned washer and it fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as the device was returned without detectable damage.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected and released to approved specifications.Device history review: a manufacturing record evaluation was performed for the finished device batch number 756d12, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N/A
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
3035526892
MDR Report Key24821934
Report Number3005075853-2026-02656
Device Sequence Number6012387
Product Code GDW
UDI-Device Identifier10705036003502
UDI-Public(01)10705036003502(17)300831(10)A98V6K
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K940967
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECS29A
Device Lot NumberA98V6K
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/25/2026
Supplement Date Received by Manufacturer04/13/2026
Initial Report FDA Received Date04/08/2026
Supplement Report FDA Received Date04/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/25/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-