| Model Number |
407145 |
| Medical Device Problem Code |
No Pacing (3268)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During a generator change, the patient was programmed to d00 mode at 70 bpm with lead check set to off.After the device was removed from the pocket, a complete loss of pacing was observed.The rv lead was disconnected from the device header and connected to a new pm, then placed into the pocket, at which point pacing resumed.The physician does not believe bovie cautery made contact with the device.No other adverse patient events were reported.Should additional information be received, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
|
| |
|