MAUDE Adverse Event Report: ABBOTT MEDICAL (CRM-SYLMAR) GALLANT¿ DR; Implantable cardioverter defibrillator (non-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL (CRM-SYLMAR) GALLANT¿ DR; Implantable cardioverter defibrillator (non-CRT) Back to Search Results
Model Number CDDRA500T
Medical Device Problem Code Failure to Convert Rhythm (1540)
Health Effect - Clinical Code Arrhythmia (1721)
Date of Event 03/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented to the emergency department with ventricular tachycardia (vt).The implantable cardioverter defibrillator (icd) and right ventricular (rv) lead failed to convert the patient¿s rhythm and external defibrillation was used.Device interrogation revealed that the rv lead had failed to capture.On (b)(6) 2026, the physician capped and replaced the rv lead.During this procedure, fluoroscopy was performed noting insulation breaks on the right atrial (ra) lead.No anomalies were noted on the rv lead via fluoroscopy.Additionally, ra lead noise was noted during manipulation.The physician elected to cap and replace the ra lead during the same procedure.The patient condition was stable before, during, and after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT¿ DR
Common Device Name
Implantable cardioverter defibrillator (non-CRT)
Manufacturer (Section D)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key24824002
Report Number2017865-2026-07315
Device Sequence Number15841875
Product Code LWS
UDI-Device Identifier05415067049476
UDI-Public(01)05415067049476(10)S000097570(17)260731
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberCDDRA500T
Device Catalogue NumberCDDRA500T
Device Lot NumberS000097570
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2026
Initial Report FDA Received Date04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age66 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-