Model Number 7120/65
Medical Device Problem Codes Failure to Capture (1081); Failure to Convert Rhythm (1540)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented to the emergency department with ventricular tachycardia (vt).The implantable cardioverter defibrillator (icd) and right ventricular (rv) lead failed to convert the patient¿s rhythm and external defibrillation was used.Device interrogation revealed that the rv lead had failed to capture.On (b)(6) 2026, the physician capped and replaced the rv lead.During this procedure, fluoroscopy was performed noting insulation breaks on the right atrial (ra) lead.No anomalies were noted on the rv lead via fluoroscopy.Additionally, ra lead noise was noted during manipulation.The physician elected to cap and replace the ra lead during the same procedure.The patient condition was stable before, during, and after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION
Common Device Name
Implantable cardioverter defibrillator (non-CRT)
Manufacturer (Section D)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key24824007
Report Number2017865-2026-07316
Device Sequence Number21221500
Product Code LWS
UDI-Device Identifier05414734502092
UDI-Public(01)05414734502092(10)2748448(17)111231
Combination Product (Y/N)Y
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2011
Device Model Number7120/65
Device Lot Number2748448
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2026
Initial Report FDA Received Date04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/11/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening;
Patient Age66 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-