• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD Back to Search Results
Model Number 0296
Medical Device Problem Codes High impedance (1291); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead was explanted due to an advisory.This rv lead was successfully explanted and replaced with another lead.No additional adverse patient effects were reported outside the revision procedure.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead was explanted due to an advisory.This rv lead was successfully explanted and replaced with another lead.No additional adverse patient effects were reported outside the revision procedure.Additional information was received that this rv lead exhibited high shock impedance recorded at approximately 115 ohms.Calcification was suspected.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Follow up report 1 (additional information): this report is being submitted to update field b5: describe event or problem, section b and field h6: device codes, section h.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24824422
Report Number2124215-2026-19210
Device Sequence Number18438402
Product Code NVY
UDI-Device Identifier00802526531279
UDI-Public(01)00802526531279(17)151028
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/28/2015
Device Model Number0296
Device Catalogue Number0296
Device Lot Number124629
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2026
Supplement Date Received by Manufacturer04/20/2026
Initial Report FDA Received Date04/08/2026
Supplement Report FDA Received Date04/24/2026
Date Device Manufactured10/28/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
Patient SexMale
-
-