• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG PROTEGO SD 65/18; ICD LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG PROTEGO SD 65/18; ICD LEAD Back to Search Results
Model Number 399410
Medical Device Problem Code Ambient Noise Problem (2877)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Biotronik will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
This lead was capped due to multiple episodes noted on recorded episodes showing noise on icd lead.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
Combination product: yes.The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGO SD 65/18
Common Device Name
ICD LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key24829931
Report Number1028232-2026-01707
Device Sequence Number13058147
Product Code NVY
UDI-Device Identifier04035479136116
UDI-Public(01)04035479136116(17)160630
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2016
Device Model Number399410
Device Catalogue Number399410
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2026
Supplement Date Received by Manufacturer04/14/2026
Initial Report FDA Received Date04/09/2026
Supplement Report FDA Received Date04/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/02/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age56 YR
Patient SexMale
-
-