|
The device was not returned for analysis.The analysis is therefore based on the received device data, as well as the inspection of the quality documents associated with the manufacture of this particular device.The manufacturing process for this device was reviewed and all production steps were performed accordingly.There was no sign of any inconsistency during the manufacturing process.The final acceptance test proved the device functions to be as specified.The returned device data was analyzed thoroughly.The analysis revealed that the clinical observation resulted from invalid memory content.The battery is not depleted.In conclusion, the device was not returned for analysis.The review of the quality documents confirmed a regular device manufacturing.The clinical observation resulted from invalid memory content.The root cause for the invalid memory content was not determinable.However, external influences, such as strong electromagnetic fields, should be taken into consideration.
|