| Model Number |
WA2T470A |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the rigid endoscope telescope had a chipped distal end.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|