D10: 300-10-15 - equinoxe replicator plate 1.5mm o/s (b)(6).300-10-45 - equinoxe replicator plate 4.5mm o/s (b)(6).300-20-02 - equinox square torque define screw drive kit (b)(6).300-30-07 - equinoxe preserve stem 7mm (b)(6).310-01-41 - equinoxe, humeral head short, 41mm (alpha) (b)(6).315-35-00 - glnd kwire (b)(6).No device was returned for evaluation; radiograph images were provided; however, the reported event was unable to be confirmed.The review identified the inferior peripheral pegs appear to have migrated from their intended positions, consistent with fracture of the glenoid component.Additionally, the center cage no longer appears parallel to the superior peripheral peg and appears that the center cage is contacting the humeral head.This appears consistent with fracture of the center cage from the glenoid body.In addition, there appears to be bone loss around the glenoid component, which may have further compromised fixation and contributed to migration of the peripheral pegs and failure of the center cage.However, the series of events cannot be confirmed as initial post-operative radiographs were not provided for evaluation.Further, the humeral head appears to be sitting proud relative to the humeral bone.According to the preserve stem operative technique, the preserve stem should be impacted until the superior face of the stem is flush with the resected humeral surface.Based on the available information, it is unclear whether the stem was initially implanted in a proud position or whether it appears proud due to observed humeral bone loss.Regardless, a proud humeral component may have further contributed to joint overstuffing, increasing eccentric loading and ultimately contributing to fracture and migration of the glenoid component.However, the sequence of events cannot be confirmed as initial post-operative radiographs were not provided.Additionally, the humeral component appears to have migrated superiorly relative to the glenoid.Post-total shoulder arthroplasty rotator cuff tearing or dysfunction is associated with proximal migration of the humeral component, which can accelerate polyethylene wear and glenoid loosening through the rocking-horse phenomenon.Although a rotator cuff deficiency was not originally reported, overstuffing of the joint is known to contribute to anatomic shoulder complications, including rotator cuff pathology, glenoid erosion, and loosening.The combination of joint overstuffing and superior humeral migration likely contributed to center cage fracture, glenoid body fracture, and subsequent migration of the glenoid component.Operative notes and/or medical records were not provided for review of usage/ technique.A review of manufacturing data was performed and no discrepancies were found.A review of complaint history determined that no further action is required as no trends were identified.A definitive root cause was unable to be determined; however, was likely associated with prosthesis wear, prosthesis fracture, and migration of the glenoid component.A combination of joint overstuffing, non-cemented use of the glenoid during implantation, and superior migration of the humeral components may have resulted in abnormal forces and eccentric loading of the glenoid, contributing to the fracture and subsequent migration.Unintended articulation between the humeral head and center cage likely resulted in the observed metal debris.However, the sequence of events cannot be confirmed, and the potential contribution of manufacturing, user, or patient-related factors could not be assessed, as the devices were not returned for evaluation and initial post-operative radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|