| Catalog Number |
2H12LP |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unk procedure, the inside of the packaging appeared cloudy, as if there was moisture.Although it seemed sterile, there were concerns, and a replacement was required.The issue was found during picking.The device was not used for the patient.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/9/2026.B3: exact event date unk, entered 1/1/2026 as only the year was provided.D4: batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/12/2026.Investigation summary: the product was returned to ees for evaluation.Visual inspection was conducted on the returned packaging.Visual inspection of the returned samples confirmed that the 2h12lp device was received within their original packaging unopened.A visual examination of the external packaging was performed, and no anomalies, damage, or irregularities were observed.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.Device history review: a manufacturing record evaluation was performed for the finished device lot 654d86 number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|