DEPUY SYNTHES PRODUCTS LLC EMAX2PLUS; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Catalog Number |
EMAX2PLUS |
| Medical Device Problem Codes |
Vibration (1674); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the electric handpiece device had vibration.It was noted in the service order that the device did not move smoothly.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: 2.3.1 visual assessment 2.4.2 loctite assessment 2.5.3 cable assessment 2.16.1 noise assessment during the service evaluation the following failures were identified: visual : cord damaged visual : breaded stainless steel wire tether damage/came off functional : noise - excessive functional : loose functional : moving parts do not move smoothly functional : vibration conclusion: failure reported is confirmed root cause: component failure from wear.
|
| |
|
Search Alerts/Recalls
|
|
|