• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KERI MEDICAL SA TOUCH® - CMC1 PROSTHESIS; TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 6MM - S

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KERI MEDICAL SA TOUCH® - CMC1 PROSTHESIS; TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 6MM - S Back to Search Results
Model Number TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 6MM - S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 02/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event could be confirmed since the affected device was returned for investigation.A review of the device history record for the reported lot did not indicate any abnormalities related to the reported event.After a thorough investigation (medical imaging, returned device, device history record review, complaint history review, surgeon's assessment), the failure appears to be primarily related to a user error (patient activity and prosthetic overstress) rather than a device malfunction.The identified root cause represents the most likely explanation based on current evidence and does not constitute definitive proof of causality.
 
Event or Problem Description
It was reported that a 48-year-old male patient (heating engineer) underwent a revision surgery due to painful crackling during movement and when using the thumb, nearly two years post initial implantation ((b)(6) 2023).During the revision surgery, the pe liner was found fractured in 2 parts.The surgeon replaced the short offset neck with a straight one.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOUCH® - CMC1 PROSTHESIS
Common Device Name
TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 6MM - S
Manufacturer (Section D)
KERI MEDICAL SA
chemin du pre-fleuri 5
plan-les-ouates, 1228
SZ  1228
Manufacturer (Section G)
KERI MEDICAL FRANCE
34 rue antoine redier
archamps, 74160
FR   74160
Manufacturer Contact
camille miletto
chemin du pre-fleuri 5
plan-les-ouates, 1228
SZ   1228
MDR Report Key24831490
Report Number3024706750-2026-00034
Device Sequence Number15742886
Product Code SFA
UDI-Device Identifier07640181160044
UDI-Public07640181160044
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P240020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberTOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 6MM - S
Device Catalogue NumberNTO156
Device Lot NumberX6366
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2025
Initial Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexMale
-
-