A device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.A trilogy 100 ventilator was returned to a third-party service center for foam remediation on the device.There was no harm or injury reported.During the evaluation of the device at the third-party service center, the device failed the active exhalation purge valve open and low o2 flow test steps.The device was requested to return unrepaired to the customer.The sound abatement foam was replaced preventatively.
|