• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L. DE CV EVOLUT PRO TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L. DE CV EVOLUT PRO TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVOLUTPRO-29-US
Medical Device Problem Codes Material Invagination (1336); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d10:  product id envpro-16-us (0012976353); product type: 0195-heart valves; select patient information cannot be included in the regulatory report due to regional privacy regulations.  medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a transcatheter aortic valve implant procedure, the first valve failed to unfold properly and did not meet the implantation requirements.Subsequently, the valve was recaptured and withdrawn from the patient.A new valve was then prepared and the delivery catheter system (dcs) was inserted into the patient.While attempting to implant the second valve, pericardial effusion was noted.In response, the valve and dcs were withdrawn from the patient.The patient was then provided surgical treatment for the pericardial effusion.
 
Additional Manufacturer Narrative
Updated data: b5.H6.Additional codes.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a transcatheter aortic valve implant procedure, the first valve failed to unfold properly and did not meet the implantation requirements.Subsequently, the valve was recaptured and withdrawn from the patient.A new valve was then prepared and the delivery catheter system (dcs) was inserted into the patient.While attempting to implant the second valve, pericardial effusion was noted.In response, the valve and dcs were withdrawn from the patient.The patient was then provided surgical treatment for the pericardial effusion.Additional information was received to clarify that the first valve unfolding was an infold that occurred after partial recapture with high positioning.The same dcs was used for the attempted implant of the first and second valves.The pericardial effusion occurred during the attempted implant of the second valve.Per the physician, the non-medtronic guidewire contributed to the pericardial effusion; however, the valve and dcs did not cause or contribute to the pericardial effusion.
 
Additional Manufacturer Narrative
Updated data: b5.Additional codes.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a transcatheter aortic valve implant procedure, the first valve failed to unfold properly and did not meet the implantation requirements.Subsequently, the valve was recaptured and withdrawn from the patient.A new valve was then prepared and the delivery catheter system (dcs) was inserted into the patient.While attempting to implant the second valve, pericardial effusion was noted.In response, the valve and dcs were withdrawn from the patient.The patient was then provided surgical treatment for the pericardial effusion.Additional information was received to clarify that the first valve unfolding was an infold that occurred after partial recapture with high positioning.The same dcs was used for the attempted implant of the first and second valves.The pericardial effusion occurred during the attempted implant of the second valve.Per the physician, the non-medtronic guidewire contributed to the pericardial effusion; however, the valve and dcs did not cause or contribute to the pericardial effusion.Additional information was received that there was no misload identified during loading of the first valve.The first valve was recaptured two times due to high positioning.A procedural delay occurred as a result of the infold.The second valve was never implanted in the patient, and the procedure was aborted due to the surgical treatment.Per the physician, the patient's severe calcification of the bicuspid aortic valve contributed to the infold.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO TRANSCATHETER AORTIC VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key24831562
Report Number9617601-2026-02366
Device Sequence Number13053624
Product Code NPT
UDI-Device Identifier00763000017835
UDI-Public(01)00763000017835(17)270620
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEVOLUTPRO-29-US
Device Catalogue NumberEVOLUTPRO-29-US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2026
Supplement Date Received by Manufacturer04/15/2026
04/21/2026
Initial Report FDA Received Date04/09/2026
Supplement Report FDA Received Date04/16/2026
04/22/2026
Date Device Manufactured08/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Patient Age67 YR
Patient SexMale
Patient Weight75 KG
-
-