SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
74027262 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, after a tka surgery, it was noticed that a jrny ii bcs xlpe art isrt sz 5-6 lt 10mm was implanted onto a size 7 jrny ii tibial base plate.The surgeon has been made aware of the situation.A revision surgery is planned to correct the implanted insert; however, the scheduled date is currently unknown.The patient¿s current health status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference number: (b)(4).The fields h3, h11 were updated.H6: results of investigation:
given the nature of the reported incident, the device could not be returned for evaluation; therefore, a device analysis could not be performed.However, according to the information provided an inappropriate implant size was used during the procedure, which necessitated the planning of a revision surgery.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for journey ii bi-cruciate stabilized (bcs) total knee system revealed in warnings and precautions that the surgeon should be familiar with the appropriate surgical techniques prior to use of this device and an adequate inventory of implant sizes should be available at the time of surgery.Additionally, revealed that the correct selection of the implant is extremely important.Failure to use the optimum size component may result in loosening, bending, cracking, or fracture of the component and/or bone.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.Device mishandling or misuse such as using the improper/incorrect procedure ( implanting mismatched sized inserts and baseplates) are documented within the complaint management system to facilitate failure monitoring and trending.Based on the nature of the failure reported and the medium risk of the complaint, no further investigation is required nor updates to instruction for use and surgical technique.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include surgical technique used or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
| |
|