Catalog Number 74027262
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/20/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, after a tka surgery, it was noticed that a jrny ii bcs xlpe art isrt sz 5-6 lt 10mm was implanted onto a size 7 jrny ii tibial base plate.The surgeon has been made aware of the situation.A revision surgery is planned to correct the implanted insert; however, the scheduled date is currently unknown.The patient¿s current health status is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference number: (b)(4).The fields h3, h11 were updated.H6: results of investigation: given the nature of the reported incident, the device could not be returned for evaluation; therefore, a device analysis could not be performed.However, according to the information provided an inappropriate implant size was used during the procedure, which necessitated the planning of a revision surgery.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for journey ii bi-cruciate stabilized (bcs) total knee system revealed in warnings and precautions that the surgeon should be familiar with the appropriate surgical techniques prior to use of this device and an adequate inventory of implant sizes should be available at the time of surgery.Additionally, revealed that the correct selection of the implant is extremely important.Failure to use the optimum size component may result in loosening, bending, cracking, or fracture of the component and/or bone.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.Device mishandling or misuse such as using the improper/incorrect procedure ( implanting mismatched sized inserts and baseplates) are documented within the complaint management system to facilitate failure monitoring and trending.Based on the nature of the failure reported and the medium risk of the complaint, no further investigation is required nor updates to instruction for use and surgical technique.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include surgical technique used or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24831564
Report Number1020279-2026-00492
Device Sequence Number9169013
Product Code JWH
UDI-Device Identifier00885556178010
UDI-Public885556178010
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74027262
Device Lot Number20CM16756
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2026
Supplement Date Received by Manufacturer05/29/2026
Initial Report FDA Received Date04/09/2026
Supplement Report FDA Received Date06/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-