| Model Number |
BF-1TH190 |
| Medical Device Problem Codes |
Melted (1385); Communication or Transmission Problem (2896); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the bronchovideoscope had a connection error during reprocessing.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, d9, h6, h11.The device was returned to olympus for inspection.In addition to the connection error, the following reportable malfunctions were identified during the device evaluation: a burnt mark in the plastic distal end cover.Based on the results of the investigation and historical data, reasonably known information suggests probable causes such as electrical problems such as open circuit, short circuit, or signal loss, without any design or manufacturing issue.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|