• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSTITUT STRAUMANN AG BL, Ø 4.8 MM RC, SLACTIVE 12 MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSTITUT STRAUMANN AG BL, Ø 4.8 MM RC, SLACTIVE 12 MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 021.6112
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could not be verified due to incomplete or missing information and / or product from the customer.Our manufacturing q-system assures that production and process controls are in place to ensure that batches confirm to the applicable specifications before theyare distributed.The loss of an endosseous dental implant after successful osseointegration and restoration is a known inherent risk of the procedure.Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.2.Absence of a recurrent peri-implant infection with suppuration.3.Absence of implant mobility.4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2011 in the patient's mouth.On (b)(6) 2024, loss of osseointegration was verified.The device was forwarded to the manufacturer.There were no patient operative or post-operative complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BL, Ø 4.8 MM RC, SLACTIVE 12 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key24833964
Report Number0009613348-2026-005316
Device Sequence Number13053516
Product Code DZE
UDI-Device Identifier07630031707215
UDI-Public07630031707215
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K062129
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number021.6112
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/09/2026
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-