Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.A2: age and date of birth unknown / not provided.A4: weight unknown / not provided.E1: email address and first/given name unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
|
It was reported that the implant at tooth site #15 was removed due to peri-implantitis.The implant was placed, and after the prosthetic restoration, about a year later, the implant came out.Symptoms as a result of the event: inflammation.
|