| Model Number |
27001 |
| Medical Device Problem Codes |
Battery Problem (2885); Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to resmed that an astral device displayed battery error messages.There was no harm or serious injury reported as a result of this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
The astral device was returned to resmed.The evaluation determined that there was no fault found with the returned device.The device was performing to specifications.The device was serviced and fully tested before it was returned to the customer.Resmed reference#: (b)(4).
|
| |
|
Search Alerts/Recalls
|