| Model Number |
97021HS |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Chest Pain (1776); High Blood Pressure/ Hypertension (1908)
|
| Date of Event |
03/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation conclusion: the customer's complaint of discrepant results with tni was not replicated with in-house retain testing of triage cardiac lot t16279rn with native whole blood.No issues with tni recovery were observed.Manufacturing batch records for the lot were reviewed and found that the lot met release specifications.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
|
| |
|
Event or Problem Description
|
|
Event occurred in the united states of america.Customer reported discrepant high tni results with multiple patients using triage cardiac lot t16279rn.Patient information: (b)(6) 2026.Sx/dx: high blood pressure and chest sensation, treatment: sent home.Medical interventions/procedures: ekgs heart score.No delay in treatment.
|
| |
|
Search Alerts/Recalls
|