| Catalog Number |
05.001.080 |
| Medical Device Problem Code |
Fail-Safe Did Not Operate (4046)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the air pen drive device run in the locked position.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: general condition.Check handpiece in ¿lock¿ mode.Check general function with apd hand switch.Check power with power test bench.Check speed with tachometer.Check for internal leak tightness.Check for external leak tightness.During the service evaluation the following failures were identified: functional: insufficient/low power.Functional: device runs in locked position.Functional: leak - air.Visual: component damaged.Conclusion: failure reported is confirmed.Root cause: faulty parts (3210).Action: repair.
|
| |
|
Search Alerts/Recalls
|