Model Number CNW0T4
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that after the intraocular lens (iol) implantation surgery, the lens was rotated to 30 or 40 degree on post operative day 1.The lenses was implanted on high myope patients with vertical lens placement.In 2 of 3 cases a capsular tension ring was used during the surgery to reduce the rotation of the lens in these patients however, the lens was not rotated in one eye and it was rotated in other eyes.Additional information has been requested, yet no new information received.There are three files associated with this report.This is 2 of 3.
 
Additional Manufacturer Narrative
The product was not returned for analysis; the lens remains implanted.Complaint history and product history record could not be reviewed because the reporting facility did not provide a valid lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON TORIC IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24839816
Report Number9612169-2026-00835
Device Sequence Number6026986
Product Code MJP
UDI-Device Identifier00380652410007
UDI-Public(01)00380652410007(17)280302(11)240304(10)25812167
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNW0T4
Device Lot Number25812167
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/19/2026
Supplement Date Received by Manufacturer05/22/2026
Initial Report FDA Received Date04/10/2026
Supplement Report FDA Received Date06/15/2026
Date Device Manufactured03/04/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REFORM CAPSULAR T RING
Patient Age59 YR
Patient SexFemale
-
-