| Model Number |
72404438 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Model number/catalo.G number: 720185-01.Serial number: n/a batch/lot number: 1100710699.Model/catalog description: reservoir flat iz 100 ml.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Incomplete inflation is acknowledged within the dfu and are not related to off-label use or failure to follow instructions.A risk review was completed, and inflation issue is defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this inflatable penile prosthesis (ipp) was having difficulty pumping the device.Troubleshooting was performed, and the physician was unsuccessful in inflating the cylinders.A surgical procedure was performed to remove and replace the ipp.No patient complications were reported.
|
| |
|