(b)(4).Date sent: 4/10/2026.D4: batch # 703d20.Investigation summary: the product was returned to ethicon endo-surgery for thorough evaluation, which included visual inspections and functional testing of the returned reload.Visual analysis of the returned sample determined that the trt10 reload was received with a damage on the distal end on the bottom side however, this damage did not have a functional impact on the reload.The reload has the proximal 21 drivers up without staples and the remaining drivers down with staples present.The returned reload had the swing tab in the locked position.In addition, an opened package was received along with the device.The reload swing tab was reset in order to fire the cartridge.The reload was tested for functionality with a test device and fired without any difficulties.The staple line was complete, and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the reload.It should be noted that the cartridge reload is designed to lockout, as a safety feature, if any staples have been fired from the cartridge reload.In addition, failure to complete the stroke may result in an incomplete staple line.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Device history review: a manufacturing record evaluation was performed for the finished device batch number 703d20, and no non-conformances related to the reported complaint condition were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|