Model Number M00550340
Medical Device Problem Codes Inflation Problem (1310); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was attempted to be used in the left main bronchus during a bronchoscopic bronchiectasis procedure for stenosis performed on (b)(6) 2026.During the procedure, the balloon leaked liquid and could not be inflated.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block e1: initial reporter facility name is (b)(6) hospital (b)(6); reported here as the facility name exceeded character limit for designated field.Block h6: imdrf device code a0504 captures the reportable event of balloon leak in the bronchus.
 
Additional Manufacturer Narrative
Block e1: initial reporter facility name is the first affiliated hospital of guangdong pharmaceutical university; reported here as the facility name exceeded character limit for designated field.Block h6: imdrf device code a0504 captures the reportable event of balloon leak in the bronchus.Block h11: investigation results the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.With all available information, boston scientific concludes the reported event of balloon leak was not confirmed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was attempted to be used in the left main bronchus during a bronchoscopic bronchiectasis procedure for stenosis performed on (b)(6) 2026.During the procedure, the balloon leaked liquid and could not be inflated.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PULMONARY
Common Device Name
BRONCHOSCOPE ACCESSORY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key24842023
Report Number3005099803-2026-01311
Device Sequence Number9928718
Product Code KTI
UDI-Device Identifier08714729456223
UDI-Public(01)08714729456223(17)280303(10)35983903
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K170759
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00550340
Device Catalogue NumberM00550340
Device Lot Number0035983903
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2026
Supplement Date Received by Manufacturer04/10/2026
Initial Report FDA Received Date04/10/2026
Supplement Report FDA Received Date04/30/2026
Date Device Manufactured03/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
Patient Weight68 KG
-
-