| Catalog Number |
133S-MV-14-T |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "spontaneous rupture".This record is for an unknown side.Device status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, corrected and/or changed data: b.5.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "spontaneous rupture".This record is for an unknown side.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "spontaneous rupture".This record is for an unknown side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation:.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- deflation: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases, openings on suture tab and wear abrasion are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|