| Model Number |
9-PDA2-05-04 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, 5mm-4mm amplatzer duct occluder was chosen for a procedure.During procedure, the device's discs was not opening properly & there was a deformation noticed in the aorta.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, 5mm-4mm amplatzer duct occluder was chosen for a procedure.During procedure, the device's discs was not opening properly & there was a bulbous deformation noticed in the aorta.The deformed device was never released from the delivery cable while inside the patient.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.The procedure got aborted.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported discharged.
|
| |
|