| Model Number |
BF-Q170 |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation and historical data, reasonably known information suggests probable causes such as damage or deterioration of parts due to wear and tear, without any design or manufacturing issue.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the bronchovideoscope image disappeared during angulation in up/down directions due to damage on charged coupled device (ccd) unit.There was no patient involvement.
|
| |
|