| Model Number |
IN2110X15B |
| Medical Device Problem Codes |
Battery Problem (2885); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy evo device was returned to a third-party service center for service.The device was not reported to be in clinical use.The trilogy evo ventilator was returned to the third party for service.During the evaluation, an error code related to a failure to charge the battery was noted.Additionally, a ventilator inoperative code related to a failure in the flow sensor was observed.The internal battery, circuit board, and flow sensor need to be replaced to address the issue.Should further investigation provide new information that impacts the outcome of the investigation or the associated medical device reporting, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of ventilator inoperative error code is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|
Search Alerts/Recalls
|