MAUDE Adverse Event Report: DEPUY MITEK LLC US MILAGRO; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MILAGRO; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 231803
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated, and no conclusion can be drawn at this time.E1: the reporter¿s complete facility address was not provided.
 
Event or Problem Description
It was reported that during an unspecified arthroscopic procedure the milagro inscr small size 5x23 screw device pulled out.Another device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: investigation summary: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that only the implant was returned for evaluation.The anchor had signs of being pulled out from the tissue, the threads were stripped.No other anomaly could be observed.The overall complaint was confirmed as the observed condition of the milagro inscr small size 5x23 would have contributed to the complained device issue.Based on the investigation findings, the potential cause can be associated with poor bone quality.As per instructions for use insufficient quantity or quality of bone, comminuted bone surfaces, or pathologic bone conditions such as cystic change or severe osteopenia that would impair the ability of the milagro interference screw to securely fixate to the bone.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MILAGRO
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key24848041
Report Number1221934-2026-01489
Device Sequence Number5663745
Product Code MAI
UDI-Device Identifier10886705010714
UDI-Public(01)10886705010714(10)143L732(17)260430
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K103831
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number231803
Device Lot Number143L732
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2026
Supplement Date Received by Manufacturer05/14/2026
05/20/2026
Initial Report FDA Received Date04/11/2026
Supplement Report FDA Received Date05/20/2026
05/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-