|
It was reported that the patient had an untreated episode stored due to t-wave oversensing.The sensing vector was set to the secondary vector.Later, during a tachycardia episode, the device had difficulty converting the arrhythmia.The arrhythmia self-converted.Also, the high energy shock impedance was 100 ohms, which is high but within normal limits.An x-ray was done and technical services advised some testing and programming options.There were no additional adverse patient effects.The system remains implanted.
|