| Model Number |
IT1032800B |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.A trilogy 200 ventilator was returned to a third-party service center for foam remediation on the device.There was no harm or injury reported.During the evaluation of the device at a third-party service center, the device was visually inspected and found no evidence of foam degradation.However, the device failed a test step related to (active exhalation purge valve closed and active exhalation proportional valve open).The sound abatement foam was replaced preventatively.The device came only for remediation and will be returned to the customer unrepaired.Additionally, during the evaluation there were no reportable error codes found in the device event log.Dust was confirmed on the blower box.The handle o ring was damaged and needs to be replaced.The software was upgraded.
|
| |
|
Additional Manufacturer Narrative
|
|
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
|
| |
|