• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROTÉGÉ MRI IPG; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROTÉGÉ MRI IPG; SCS IPG Back to Search Results
Model Number 3771
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 03/05/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
B3: event date is estimated.During processing of this incident, attempts were made to obtain complete patient information.Further information was requested but not received.
 
Event or Problem Description
It was reported that the patient had pain at the ipg site.Surgical intervention was undertaken and the system was explanted.
 
Additional Manufacturer Narrative
The protégé ipg was explanted due to discomfort.This event is related to patient factors and cannot be confirmed through product analysis testing.Per event details, the patient had lost a lot of weight and the implant was causing discomfort.There were no allegations of deficiency against the ipg.Visual inspection of the ipg did not identify any anomalies that would contribute to the reported issue.The ipg communicated and charged normally with lab standard equipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTÉGÉ MRI IPG
Common Device Name
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
9723098000
MDR Report Key24854027
Report Number3006705815-2026-02271
Device Sequence Number10549146
Product Code LGW
UDI-Device Identifier05415067017086
UDI-Public(01)05415067017086(10)A000029699(17)181004
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/04/2018
Device Model Number3771
Device Catalogue Number3771
Device Lot NumberA000029699
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2026
Supplement Date Received by Manufacturer07/06/2026
Initial Report FDA Received Date04/13/2026
Supplement Report FDA Received Date07/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS LEAD
Outcome Attributed to Adverse Event Other;
Patient Age52 YR
Patient SexMale
-
-