MAUDE Adverse Event Report: ABBOTT MEDICAL (CRM-SYLMAR) AVEIR¿ DELIVERY CATHETER; LEADLESS PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL (CRM-SYLMAR) AVEIR¿ DELIVERY CATHETER; LEADLESS PACEMAKER Back to Search Results
Model Number LSCD111
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Syncope/Fainting (4411)
Date of Event 03/19/2026
Type of Reportable Event Serious Injury
Event or Problem Description
During the leadless pacemaker (lp) system implant procedure, due to patient's small heart anatomy, it was difficult to access the right ventricular (rv).When the lp was maneuvered into a good position in the rv, the introducer was lowered and the lp positioning was lost.At that time the patient experienced a drop in blood pressure resulting in syncope.A medication was administrated to stabilize the patient.The lp system was explanted and the physician elected to implant a traditional pacemaker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEIR¿ DELIVERY CATHETER
Common Device Name
LEADLESS PACEMAKER
Manufacturer (Section D)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key24854258
Report Number2017865-2026-07581
Device Sequence Number16089257
Product Code PNJ
UDI-Device Identifier05415067036384
UDI-Public(01)05415067036384(10)11059797(17)261130
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P150035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLSCD111
Device Catalogue NumberLSCD111
Device Lot Number11059797
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2026
Supplement Date Received by Manufacturer06/23/2026
Initial Report FDA Received Date04/13/2026
Supplement Report FDA Received Date06/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-