• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC FULL OSSEOTITE® TAPERED IMPLANT 4 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC FULL OSSEOTITE® TAPERED IMPLANT 4 X 11.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number FNT411
Medical Device Problem Codes Loss of Osseointegration (2408); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Osteolysis (2377)
Date of Event 05/15/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier: unknown / not provided.A4: patient weight: unknown / not provided.D10: concomitant medical product: fnt410, full osseotite® tapered implant 4 x 10mm, lot: 2019040870 (x4).Fnt411, full osseotite® tapered implant 4 x 11.5mm, lot: 2018100391 (x3).A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported the implants at tooth site 12,14,16,17, 22, 24, 26, 27 lost integration due to peri-implantitis and were removed.Patient referred pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL OSSEOTITE® TAPERED IMPLANT 4 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24855545
Report Number0001038806-2026-01978
Device Sequence Number10493404
Product Code DZE
UDI-Device Identifier00844868003335
UDI-Public(01)00844868003335(17)231017(10)2018100391
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K063286
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/17/2023
Device Catalogue NumberFNT411
Device Lot Number2018100391
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/17/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL IMPLANTS SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-