H11: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.D2b (neu; gdw).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
H11: as per the clarification received from the customer, the cannula was empty and there was no clip present inside the device.Based on this information, the originally reported issue coded as "detachment of device or device component" has been revised to "device slipped." there is no detachment-related failure reported, the event is considered not reportable.Accordingly, the file was reassessed for reportability and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.Manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported device slipped issue could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, d2b (neu; gdw), g3, h6 (device, component, method, result, conclusion)
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
On (b)(6) 2026, a patient underwent an ultrasound guided breast marker placement procedure through hard density tissue using ultra clip tissue marker device.During the procedure, the device clip coming loose.No coaxial needle was used.There was no reported patient injury.
|