• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH CYSTOSCOPE, 30°, 17 FR. X 225 MM, 9 FR. CHANNEL, OCULAR ANGLE 45°; RIGID CYSTOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH CYSTOSCOPE, 30°, 17 FR. X 225 MM, 9 FR. CHANNEL, OCULAR ANGLE 45°; RIGID CYSTOSCOPE Back to Search Results
Model Number A20919A
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is presumed that the reported event occurred due to component failure of the device.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the rigid scope had aurum plating distal end peeled off.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTOSCOPE, 30°, 17 FR. X 225 MM, 9 FR. CHANNEL, OCULAR ANGLE 45°
Common Device Name
RIGID CYSTOSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24855961
Report Number9610773-2026-02943
Device Sequence Number9009012
Product Code GCP
UDI-Device Identifier04042761030052
UDI-Public04042761030052
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberA20919A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/26/2026
Initial Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-