• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 2.0 25; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 2.0 25; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX643T
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 02/25/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: during the investigation, no significant deformations or damage of the valve were determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the progav 2.0 operates within the accepted tolerance in the horizontal position.The shuntassistant 2.0 does not operate within the specified tolerances in the vertical position.An accelerated outflow of shuntassistant 2.0 could be determined.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerances.Internal inspection : after dismantling of the valves, deposits were found.Results based on our investigation results, we determined visible deposits in the progav 2.0.These deposits did not affect the device's technical performance at the time of the investigation.Furthermore, an accelerated flow rate in the shuntassistant 2.0 has been observed.The deposits visible in the valve have led to a change in flow rate.When the shuntassistant is opened, it is possible to view a large portion of the interior.However, there are areas that cannot be inspected visually.Organic deposits may be present in these areas, which could impair functionality.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
 
Event or Problem Description
It was reported that a progav 2.0 shuntsytsem (#fx643t) was implanted during a procedure performed on (b)(6) 2025.According to the complainant, the shunt system caused an underdrainage.The patient underwent a revision procedure on (b)(6) 2026.The complained product was sent to the manufacturer for investigation.No patient complications were reported as a result of the revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 2.0 25
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key24860054
Report Number3004721439-2026-00092
Device Sequence Number10549827
Product Code JXG
UDI-Device Identifier04041906502010
UDI-Public(01)04041906502010(11)241223(17)291219(10)20079575
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX643T
Device Catalogue NumberFX643T
Device Lot Number20079575
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2026
Initial Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
Patient Weight85 KG
-
-