| Catalog Number |
05.001.202 |
| Medical Device Problem Codes |
Leak/Splash (1354); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: section 70: information button & self-test section 80: charging and checking of power module in charger ubc ii section 120: function ¿ test section 130: saw ¿ test section 140: check liquid indicator during the service evaluation the following failures were identified: functional: will not charge functional: led warning light indicator error - service functional: leak - liquid.Conclusion: ¿ failure reported is confirmed ¿ root cause: user error.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the power module device had a liquid leak.It was noted in the service order that the device did not work.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|