• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORPORATION QFIX 1.8 MINI SUTURE ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORPORATION QFIX 1.8 MINI SUTURE ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72290123
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 03/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a biceps tenodesis procedure, the q-fix suture detached from the anchor after the deployment knob was turned and the suture spool was unwound.Two pieces of suture with frayed ends were removed from the proximal end of the inserter.The suture anchor was not visible, and it is unclear whether the anchor remained inside the patient.The procedure was resumed, without any delay, using an equivalent s+n back-up.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference:(b)(4).H11 h3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the device, the anchor is not visible but the sutures are broken.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the suture strength requirements are specified.A certificate of analysis is required with each shipment verifying suture diameter & knot pull suture strength.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include excessive force used on the device or an impact event inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QFIX 1.8 MINI SUTURE ANCHOR
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24860655
Report Number3006524618-2026-00181
Device Sequence Number18451317
Product Code MBI
UDI-Device Identifier00885556662618
UDI-Public885556662618
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172165
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number72290123
Device Lot Number2184969
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2026
Supplement Date Received by Manufacturer06/03/2026
Initial Report FDA Received Date04/13/2026
Supplement Report FDA Received Date06/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-